A patient monitor is not chosen by screen size but by the exact list of parameters your department needs — today and in three years. An under-equipped monitor gets replaced; a monitor overpaid in unused options is capital tied up.

Parameters, by level of care

The configuration follows directly from the department:

  • Recovery, ward, consultation: 3-lead ECG, SpO2, non-invasive blood pressure (NIBP), temperature. The minimum base.
  • Emergency, transport: add 5-lead ECG, respiration, and real battery life above 3 hours.
  • Intensive care: add invasive pressure (at least 2 channels), capnography (EtCO2), two temperature channels, and cardiac output depending on the case mix.
  • Operating theatre: capnography and anaesthetic agent analysis become essential, often built into the anaesthesia machine rather than the monitor.

Modular or fixed configuration?

A fixed-configuration monitor costs less and fails less often: fewer connectors, fewer points of failure. It suits departments whose needs are stable.

A modular monitor lets you add a parameter later and move a module from one bed to another. It is justified in intensive care and in departments whose remit changes. Get the unit price of every module before taking that route: the final gap often exceeds the cost of two fixed monitors.

SpO2: the sensor question

Oximetry technology determines reliability under motion or poor perfusion. More importantly, sensors are proprietary: ask for the price of a reusable adult sensor, a paediatric sensor and a disposable one. Over five years, SpO2 consumables can exceed the price of the monitor.

The same reasoning applies to NIBP cuffs, ECG cables and capnography lines.

Alarms and safety

Alarms are a clinical subject, not a technical detail. Check for:

  • priority levels compliant with IEC 60601-1-8 (high, medium, low), audible and visual;
  • thresholds adjustable per parameter and storable per patient profile;
  • a minimum volume that cannot be disabled;
  • screen legibility at 3 or 4 metres, standing, from the door of the room.

Networking and the patient record

Above three beds, a central station changes how the department works: one person watches every patient. Put three questions to the supplier:

  • Is the central station included or charged separately, and up to how many beds?
  • Does the monitor export in HL7 to an electronic patient record?
  • How much trend data does the device hold internally (24 h, 72 h, 96 h)?

Battery life and robustness

Quoted battery life is measured under optimal conditions. Ask for the figure with automatic NIBP every 15 minutes and continuous SpO2 — that is the only useful number. Plan for battery replacement too: it is a wear part, changed every two to three years.

For transport use, check the ingress protection rating and shock resistance.

Your checklist

  • Which parameters are genuinely used in the department, and which are anticipated?
  • Unit price of every sensor and consumable, over 5 years?
  • Are alarms compliant and audible from the corridor?
  • Is HL7 export available? Is the central station included?
  • Battery life under real conditions?
  • Who services it, and with what local parts stock?